Instruction Manual
Model 755A
Electronic circuitry and internal interconnection wiring is shown in the schematic diagrams and wiring diagram in the rear of this manual. For detailed circuit analysis, refer to Section 5.
a.Detector/Magnet Assembly
A
The difference between the signals developed by the two halves of the photocell constitutes the error signal supplied to the input of amplifier U1 on the Control Board. Amplifier U1 drives U2 which, in turn, supplies the restoring current to the titanium wire loop on the test body (see Section
Elements of Detector Temperature Control Circuit
Detector temperature is sensed by thermistor RT1, an integral part of the detector assembly (see Figure
The output from this circuit is applied to the two heaters within the detector/magnet assembly; HR1, mounted on the top of the magnet, and HR2, mounted permanently on the rear of the detector assembly.
b.Control Board and Associated Circuitry
The Control Board contains signal conditioning and control circuitry. The Control Board is mounted on the inside of
the analyzer door (see Figure
The Control Board contains the following:
Input Amplifier U1
This amplifier receives the error signal from the dual photocell of the detector assembly and drives amplifier U2.
Amplifier U2 and Associated Zero Adjustment
Amplifier U2 supplies the restoring current to the titanium wire loop of the test body within the detector assembly. Front panel ZERO Control R13 applies an adjustable zero biasing signal to the input of U2 to permit establishing a zero calibration point on the display or recorder. With downscale zero standard gas flowing through the analyzer, the ZERO control is adjusted for the appropriate reading.
Amplifier U4 and ASsociated Span Adjustment
Amplifier U4 and associated feedback resistors provide a signal amplification of X4, resulting in a signal level suitable for analog divider circuit U6. Front panel SPAN adjustment R16 modifies the value of the input resistance and hence the signal amplification factor. Adjustment range is approximately ±30%.
The SPAN adjustment permits establishing an upscale calibration point on the display or recorder. With upscale standard gas flowing through the analyzer, the SPAN control is adjusted for the appropriate reading.
Pressure Compensation Circuit
The pressure compensation circuit consists of divider U6 and associated components. This circuit provides a
Vo = k (Vx/Vz)
Theory | Rosemount Analytical Inc. A Division of Emerson Process Management |