Model 755A
Instruction Manual
Variables that influence paramagnetic oxygen measurements include: operating pressure, Section
a.Pressure Effects
Although normally calibrated for readout in percent oxygen, the Model 755A actually responds to oxygen partial pressure. The partial pressure of the oxygen component in a gas mixture is proportional to the total pressure of the mixture. Thus readout is affected by pressure variations. However, the instrument incorporates electronic circuitry that provides automatic pressure compensation for pressure variations within ±3% of the target value, which may be set anywhere in the range of
WARNING
POSSIBLE EXPLOSION HAZARD
Never subject the sensing unit to an absolute pressure of less than 500 mm Hg (66.7 kPa).
b.Temperature Effects
Magnetic susceptibilities and partial pressures of gases vary with temperature. In the Model 755A,
1.Interior of the analyzer case is maintained at 140°F (60°C) by an electronically controlled heater and associated fan.
2.Immediately downstream from the inlet port, prior to entry into the detector, the sample is preheated by
passing through a coil maintained at approximately the same temperature as the detector.
3.The detector is maintained at a controlled temperature of 150°F (66°C).
c.Interferents
Instrument response to most
During initial installation of an instrument for a given application, effects of the background gas should be calculated to determine if correction is required. See Section
d.Vibration Effects
Instrument Design
To minimize vibration effects, the detector/magnet assembly is contained in a
Installation
Avoid excessive vibration. In making electrical connections, do not allow any cable to touch the
Electronic Response Time
If readout is noisy despite precautions mentioned above, obtain slower electronic response by counterclockwise adjustment of Response Time Adjust R30 (see Table
Rosemount Analytical Inc. A Division of Emerson Process Management | Theory |