Instruction Manual
Model 755A
| |
CHANGES ON INSTRUMENT READOUT |
After the calibration procedure of Section
At this time, an adjustment of electronic response time (R30 on the Control Board, see Table
If the analyzer exhaust port is vented through a suitable absolute backpressure regulator, and the installation site is at or near sea level, barometric pressure changes do not affect the percent oxygen readout. If conditions are otherwise, the pressure compensation circuit must be
The appropriate calibration interval will depend on the accuracy required in the particular application, and is best determined by keeping a calibration log.
Operation | Rosemount Analytical Inc. A Division of Emerson Process Management |