Preparatory to startup and calibration, the operator should study Figure
The digital display provides readout over the full range of 0.00% to 100.00% oxygen, eliminating the requirement for expanded- scale operation. With a recorder, however, resolution is normally limited to approximately 1% of fullscale. Thus, if the recorder output is to be the important display medium for the particular application,
a.Recorder Oxygen Range Selection Procedure
3.Select the required zero suppression by appropriate settings of the following:
a.Set Recorder Zero Suppression Jumper for fixed value of 0%, 20%, 40%, 60% or 80% oxygen.
b.Set Recorder Zero Suppression Coarse Adjustment R41 for appropriate value in the range of 05 to 25% oxygen.
NOTE:
The actual applied zero suppression may be measured and established via readout on the digital display per the procedure of Section
Example 1, Selection of a
•Desired oxygen range for recorder output: 0 to 100%.
Refer to Table
1.Verify that Recorder Voltage Output Jumper is in the position appropriate to the recorder: 10 mV, 100 mV, 1 V or 5 VDC.
2.Place the Recorder Oxygen Span Jumper in the position appropriate to the desired span. Note that on the circuit board switch positions are marked according to the amplifier gain.
•Required span of 100% oxygen is selected when Recorder Oxygen Span Jumper is in the 1X position.
•Required Zero Suppression is 0%, thus Recorder Zero Suppression Jumper, is removed, and R41, Recorder Zero Suppression Coarse Adjustment, and R104, Recorder Zero Suppression Fine Adjustment, are adjusted for 0% oxygen.
Desired Oxygen | Amplifier Gain |
Span for Recorder | Marking on Board |
100% | 1X |
20% | 5X |
10% | 10X |
5% | 20X |
2% | 50X |
1% | 100X |