Instruction Manual
Model 755A
If the lamp assembly removed from the instrument has four wires, the 633689 Connector Board requires modification. Continue to step 10.
10.Refer to Figure
11.Place Connector Board on a clean working surface, with solder side (no components) up.
12.Per Figure
NOTE
If the Connector Board cannot be satisfactorily modified, a modified 633689 Connector Board may be ordered from the factory. See Section 7.
13.Reassemble detector, etc., in reverse order of disassembly.
14.Refer to Figure
15.Realign the photocell per Section
Add Straps or Solder Bridges
F3
HR1
Solder Side of Board (Backside)
Figure
1/4"
Set Screw
Red Mark for
Alignment
Figure
Service and Maintenance | Rosemount Analytical Inc. A Division of Emerson Process Management |