Instruction Manual

760004-A February 2002

Model NGA 2000 HFID

1-5 SPECIFICATIONS

a.

General

 

 

Measurement Species ....................

Total hydrocarbons

 

Ranges (H2/He Fuel)

 

 

Low range..............................

0 to 10 ppm, CH4, through 0 to 1%, CH4 at an oven setpoint be-

 

 

tween 113°C and 191°C

 

High range .............................

0 to 50 ppm, CH4, through 0 to <5%, CH4 at an oven setpoint be-

 

 

tween 113°C and 191°C

 

Analysis Temperature .....................

Adjustable from 200oF to 400oF (93oC to 204oC), maintained within

 

 

±11oF (±6oC) from the setpoint.

 

Repeatability ....................................

1% of fullscale for successive identical samples, at a constant

 

 

temperature, sample flow and fuel, burner air, regulated air and

 

 

sample pressures

 

Min. Detectable Level......................

0.10 ppm, CH4

 

Noise ................................................

1% of fullscale, peak to peak

 

Linearity............................................

±1% of fullscale, ±2% of data point (must be above the minimum

 

 

detectable level)

 

Response Time ...............................

1.5 sec., 0% to 90% of fullscale

 

Drift

 

 

Zero .......................................

±1% of fullscale/24 hours at constant temperature, sample flow,

 

 

hydrocarbon concentration of supply gases, and

 

 

fuel, burner air, regulated air and sample pres-

 

Span ......................................

sures.

 

±1% of fullscale/24 hours at constant temperature, sample flow,

 

 

hydrocarbon concentration of supply gases, and

 

 

fuel, burner air, regulated air and sample pres-

 

 

sures.

 

Effect of Temperature....................

±2% of fullscale for any ambient temperature change of 10oC and

 

 

rate of change less than 10oC/hr.

 

Operating Temperature .................

59°F to 95°F (15°C to 35°C)

 

Power Requirements.......................

+24 VDC ±5%, 120 W max. direct to Analyzer Module

 

 

Ripple and Noise: <100 mV pp

 

 

Line and Load Regulations: <1%

b.

Physical

 

 

Case Classification........................

General purpose for installation in weather-protected area

 

Maximum Separation ....................

1600m (1 mile) from Analyzer Module to Platform

 

Materials in Contact With Sample . Stainless steel and glass-filled Teflon

 

Dimensions....................................

See Outline and Mounting Dimensions, Figure 2-5

 

Weight ...........................................

15.9 kg (35 lbs.)

 

Mounting........................................

Horizontally, custom-installed in a panel

1-4 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management