Instruction Manual
Model NGA 2000 HFID
a. | General |
|
| Measurement Species .................... | Total hydrocarbons |
| Ranges (H2/He Fuel) |
|
| Low range.............................. | 0 to 10 ppm, CH4, through 0 to 1%, CH4 at an oven setpoint be- |
|
| tween 113°C and 191°C |
| High range ............................. | 0 to 50 ppm, CH4, through 0 to <5%, CH4 at an oven setpoint be- |
|
| tween 113°C and 191°C |
| Analysis Temperature ..................... | Adjustable from 200oF to 400oF (93oC to 204oC), maintained within |
|
| ±11oF (±6oC) from the setpoint. |
| Repeatability .................................... | ≤1% of fullscale for successive identical samples, at a constant |
|
| temperature, sample flow and fuel, burner air, regulated air and |
|
| sample pressures |
| Min. Detectable Level...................... | 0.10 ppm, CH4 |
| Noise ................................................ | ≤1% of fullscale, peak to peak |
| Linearity............................................ | ≤±1% of fullscale, ≤±2% of data point (must be above the minimum |
|
| detectable level) |
| Response Time ............................... | ≤1.5 sec., 0% to 90% of fullscale |
| Drift |
|
| Zero ....................................... | ≤ ±1% of fullscale/24 hours at constant temperature, sample flow, |
|
| hydrocarbon concentration of supply gases, and |
|
| fuel, burner air, regulated air and sample pres- |
| Span ...................................... | sures. |
| ≤ ±1% of fullscale/24 hours at constant temperature, sample flow, | |
|
| hydrocarbon concentration of supply gases, and |
|
| fuel, burner air, regulated air and sample pres- |
|
| sures. |
| Effect of Temperature.................... | ≤±2% of fullscale for any ambient temperature change of 10oC and |
|
| rate of change less than 10oC/hr. |
| Operating Temperature ................. | 59°F to 95°F (15°C to 35°C) |
| Power Requirements....................... | +24 VDC ±5%, 120 W max. direct to Analyzer Module |
|
| Ripple and Noise: <100 mV pp |
|
| Line and Load Regulations: <1% |
b. | Physical |
|
| Case Classification........................ | General purpose for installation in |
| Maximum Separation .................... | 1600m (1 mile) from Analyzer Module to Platform |
| Materials in Contact With Sample . Stainless steel and | |
| Dimensions.................................... | See Outline and Mounting Dimensions, Figure |
| Weight ........................................... | 15.9 kg (35 lbs.) |
| Mounting........................................ | Horizontally, |
Rosemount Analytical Inc. A Division of Emerson Process Management |