Model NGA 2000 HFID

Instruction Manual

760004-A February 2002

SECTION 3

OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should leak test the module as outlined in Section 2.

For the remainder of this section, Analyzer Module interconnection with a control module or some interfacing component will be assumed operational.

3-2 STARTUP PROCEDURE

WARNING

PRESSURIZED ENCLOSURE

This enclosure shall not be opened unless the area is known to be free of flammable materials or unless all devices within have been de-energized.

Area classification for the protected enclo- sure:Non-Classified.

Pressurization: Type Z

Temperature Identification Number: T4A

Power shall not be restored after enclosure has been opened (or loss of purge) until enclosure has been purged for a minimum of 6 minutes at the minimum pressure of 689 hPa (10 psig). For safety, the Analyzer Module should be installed in a non-confined, ventilated space. Do not block any of the rear panel outlets as they are part of the safety system.

1.Connect supply gases and outlets to/from module.

2.Turn ON the purge gas only. Perform a leak check. Wait a minimum of 6 min- utes.

3.Connect the LON cable(s) and the +24VDC power cable.

4.Turn power ON.

5.Check the 4 LEDs. The power green LED should be illuminated. The Oven amber LED should be blinking or on. The other LEDs should be OFF.

6.Allow the network to initialize.

7.If the user's system contains only one Analyzer Module, all system components, the Controller Board and the network "self-install" (bind together) during initial startup. If the system contains more than one Analyzer Module, the startup se- quence will interrogate the network to lo- cate and identify all components on the network. The user will have to bind ap- propriate combinations of components after the startup sequence. (See Section 3-3.)

8.Check the general health of the analyzer by reviewing the status of the Self Tests. All “Pass” conditions should be obtained.

These test results can be found by se- lecting the following from the Main Menu: Technical Level Configuration, Diagnostic Menus, Analyzer Module Diagnostics, Self Test. All tested parameters should indicate "Pass."

Descriptions of the tests performed follow:

EEPROM test - Checks the EEPROM on the Analysis Computer PCB.

EPROM test - Checks the EPROM on the Analysis Computer PCB.

RAM test - Checks the RAM on the Analysis Computer PCB.