SECTION 4

MAINTENANCE AND SERVICE

WARNING

QUALIFIED PERSONNEL

This equipment should not be adjusted or repaired by anyone except properly qualified service personnel.

4-1 OVERVIEW

This section contains instructions and procedures for troubleshooting and maintaining the HFID analyzer module. To access the internal components of the analyzer module, perform the following:

1.Remove power to the unit; shut off gases and disconnect lines. Allow module to cool.

2.Refer to Figure 4-1. Remove the six screws securing the front panel, then the six screws securing the cover to the rear panel. Slide cover towards rear panel to remove. Loosen four screws securing in- ner insulation shield to base, lift up to re- move.

Figure 4-2 illustrates the locations of major components of the HFID.