b.Replacing Burner Jets

Disassemble the burner only if contami- nants are evident. Combustion products or other contaminates which accumulate inside the burner may form electrical leakage paths between the collector and the burner contact, resulting in noisy readings.

If the analyzer module is to be operated at the highest sensitivity, traces of such contaminates can cause erroneous read- ings. For best performance, replace the burner jet follows:

1.Remove oven from analyzer module per Section 4-2a.

2.Remove burner/thermal block from oven per Section 4-2b.3.Remove thermal block from burner per Section 4-4a.

4.Refer to Figure 4-7A. Remove screws (2) holding burner cap re- tainer, remove retainer.

5.Holding burner base, lift burner cap off of assembly, set aside, remove gasket.

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the burner with bare hands. All tools used for maintenance must be free of contaminates.

6.Refer to Figure 4-7B. Holding burner base, lift combustion chamber off, set aside.

A

Burner Cap

 

Retainer

Burner Cap

Gasket 656931

B

Combustion Chamber

O-Ring 904373

Burner Base

Burner Base

Figure 4-7. Burner Disassembly