4-2 OVEN

NOTE

Though the oven can be replaced as a complete unit, all internal components are field re- placeable.

DO NOT remove the fittings from the gas lines on the detector.
a.Removal

Refer to Figure 4-3, disconnect the oven's three gas lines and seven electrical ca- bles, noting location of mating connectors for re-installation.

Remove the two hex nuts securing the oven to the chassis and the two screws securing oven to the rear panel. Lift oven assembly from ana- lyzer.

Oven Assembly 659551

Figure 4-3. Removal of Oven from Chassis

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service

4-3