Instruction Manual

760004-A February 2002

Model NGA 2000 HFID

b.Case Temperature Sensor

1.Per Section 4-6, remove securing hardware from electronics assembly and lay on bench.

2.Disconnect case temperature sensor cable.

3.Remove screw securing cable clamp holder to signal board.

4.Remove case temperature sensor from cable clamp holder.

5.Per Figure 4-13 insert replacement case temperature sensor into cable clamp holder.

6.Re-assemble to signal board mount- ing screw.

Center sensor in cable clamp

Figure 4-13. Case Sensor Installation

4-16

Maintenance and Service