4-4 BURNER INTERNAL COMPONENTS2. Remove burner/thermal block from

 

oven per Section 4-2b.

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the burner with bare hands. All tools used for maintenance must be free of contaminates.

a.Disassembly of Burner/Thermal Block1.Remove oven from analyzer module per Section 4-2a.
3.Refer to Figure 4-6. Disconnect sam- ple capillary nut at base of burner.4.Remove screw securing thermal block to burner.5.Carefully pull burner away from ther- mal block.

Burner

Sample capillary input

Thermal Block

Sample capillary

Figure 4-6. Burner/Thermal Block Disassembly

4-8

Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management