WARNING
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
CAUTION
If this Analyzer Module is used with a
WARNING
PRESSURIZED ENCLOSURE
HOT INTERNAL PARTS
This enclosure shall not be opened unless the area is known to be free of flammable materials or unless all devices within have been
Area classification for the protected enclosure:
Nonclassified.
Pressurization: Type Z
Temperature Identification Number: T4A
Power shall not be restored after enclosure has been opened (or loss of purge) until enclosure has been purged for a minimum of 6 (six) minutes at the minimum pressure of 689 hPa (10 psig)