Model NGA2000 HFID
Instruction Manual
Noncompliance could cause damage to the module. At the very least, the mod- ule's safety system, which requires a certain volume of purge air flowing through the case before allowing burner ignition, will not allow the instrument to operate. The lowest purge air flow/pressure setting possible during burner operation is preferable. Thus, the user should set the external purge air pressure initially at 689
69
DO NOT EXCEED 1378
Note
Do not restrict the PURGE OUT port and the pressure relief valve. They must be vented to atmospheric pressure.
12.Manual or
Two methods of burner ignition are pos- sible:
Before ignition and operation, Fuel Flow must be set to ON in "Light Flame" dis- play screen under Basic Controls and oven temperature must be at least 85°C.
The manual ignition switch on the Analyzer Module front panel must be manipulated in the following ways:
•Press up and hold for one minute. This opens burner fuel and air sole- noids.
•Press down to ignite burner glow plug for up to 10 seconds.
•Repeat as necessary (if fuel and air sources are farther away than 10 feet, several more attempts may be necessary).
•If the flame has been lit, but the flame temperature increases slowly, perform the following steps:
!After igniting flame, release switch for 2 seconds
!Press switch down for 2 seconds
!Repeat release switch and press down steps as necessary.
13.Allow the case and oven to warm up, approximately 1 to 2 hours.
14.Verify that all 4 LEDs are illuminated.
Note the four LEDs on the front panel of the Analyzer Module. They provide nec- essary information for either ignition procedure. The LEDs, when illuminated, denote the following information:
•Green - unit powered on
•Amber - continuous illumination im- plies oven has reached operating temp. Within ±6°C of setpoint
•Green - Flame on
•Green - purge air system intact (it has filled five volumes of the module interior)
Rosemount Analytical Inc. A Division of Emerson Process Management | Startup and Operation |