Instruction Manual
Model NGA2000 HFID
The HFID Analyzer Module is designed with a
The purge system is enabled only if there is proper purge gas flow in, purge gas pressure, and internal case pressure, and after five times the case volume has been exchanged.
All tubing ahead of the burner is rigid metallic tubing assembled with ferrule/nut type compression fittings. However, should an internal fuel leak occur, a
dissipated below 25% of the LEL of hydrogen through the combination of an inlet fuel flow restrictor and purge gas flow.
This module is designed to use 40% H2/60% He fuel at a maximum inlet pressure of 3446
A standard HFID Analyzer Module is only equipped to analyze a
WARNING
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon a special fuel flow restrictor in the fuel inlet fitting. Do not remove fuel inlet restrictor. Do not use 100% hydrogen fuel. Replace only with a factory supplied fit- ting.
Description and Specifications | Rosemount Analytical Inc. A Division of Emerson Process Management |