Case Temperature Sensor

1.Per Section 4-2e on page 4-13, re- move securing hardware from elec- tronics assembly and lay on bench.

2.Disconnect case temperature sensor cable.

3.Remove screw securing cable clamp holder to signal board.

4.Remove case temperature sensor from cable clamp holder.

5.Per Figure 4-13 below insert re- placement case temperature sensor into cable clamp holder.

Re-assemble to signal board mounting screw.

Center

sensor in

cable clamp

Figure 4-13. Case Temperature Sensor Installation