Instruction Manual
Model NGA2000 HFID
Installation
WARNING
BURNER CONTAMINATION
Do not handle internal parts of the burner with bare hands. All tools used for maintenance must be free of con- taminates.
1.Install new lower gasket, jet assembly and upper gasket into burner base,
2.Install new sample jet (with gasket) and tighten.
3.Tighten jet nut.
4.Install air baffle per Figure
Note
Incorrect installation of air baffle will cause ignition failure.
5.See Figure
6.Set combustion chamber into burner base being careful not to move air baffle.
7.See Figure
Rosemount Analytical Inc. A Division of Emerson Process Management |