Model NGA2000 HFID
Instruction Manual
b. Gas Requirements |
|
|
|
Sample .......................................... | |||
Flow rate .............................. | 1.0 to 2.5 L/min. | ||
Supply pressure................... | 345 to 620 | ||
Temperature ........................ | 110oC to 230oC (230oF to 446oF), <20oC variance/24 hours, <10oC | ||
| variance/hr. | ||
Particulates.......................... | filtered to <2 microns | ||
Dewpoint.............................. | <15oC below the setpoint | ||
Regulated Air................................. | Instrument air or nitrogen | ||
Flow rate .............................. | 2 to 4 L/min. | ||
THC ..................................... | ≤2 ppm, CH4 | ||
Supply pressure................... | 689 to 1723 | ||
Particulates.......................... | filtered to <2 microns | ||
Purge Air:....................................... | Instrument air, nitrogen or other nonflammable gas (refer to | ||
| ANSI/NFPA 496 for the requirements for the Protective Gas Sys- | ||
| tem) | ||
Flow rate: ............................ | 16 to 18 L/min. | ||
Supply pressure:.................. | 689 to 1378 | ||
Fuel Gas ........................................ | Premixed 40% hydrogen and 60% helium | ||
Flow rate .............................. | 80 to 100 ml/min | ||
THC ..................................... | ≤0.5 ppm, CH4 | ||
Supply pressure................... | 3101 to 3446 | ||
|
|
|
|
|
|
|
|
|
| WARNING | |
| EXPLOSION HAZARD |
Do not use pure hydrogen fuel. An explosion resulting in severe personal injury or death could oc- cur.
Burner Air ...................................... | |
Flow rate .............................. | 350 to 400 mL/min. |
THC ..................................... | ≤1 ppm, CH4 |
Supply pressure................... | 1723 to 3446 |
c. Physical |
|
Case Classification........................ | General purpose for installation in |
Maximum Separation .................... | 1600m (1 mile) from Analyzer Module to Platform |
Materials in Contact With Sample . Stainless steel and | |
Dimensions.................................... | See Outline and Mounting Dimensions, Figure |
Weight ........................................... | 15.9 kg (35 lbs.) |
Mounting........................................ | Horizontally, |
Rosemount Analytical Inc. A Division of Emerson Process Management | Description and Specifications |