Model NGA2000 HFID
Instruction Manual
b.Pneumatic Connections
WARNING
HIGH TEMPERATURE
The Sample In, Sample Bypass Out, and Burner Exhaust Out gases and fittings can reach temperatures of up to 250°C.
Make connections to these fittings when the oven heater is disabled or the module is powered down.
(See Figure
It is recommended that no connection be made to the PURGE AIR OUT port. If, how- ever, the analyzer's location requires interconnection with a venting system, the 3/8" O.D. line should be kept as short as possi- ble, and no longer than four feet.
CAUTION
POSSIBLE INSTRUMENT DAMAGE
No connection should be made to the PRESSURE RELIEF VALVE fitting. Doing so may cause damage to the instrument.
CAUTION
PURGE AIR REQUIREMENTS
The front panel LEDs of the Analyzer Module, as installed, are not visible, the user should provide an indicator for the safety system as per ANSI/NFPA 496 standards.
c.Fuel Gas Specification
Standard analysis usually requires mixed fuel, i.e., 40% ±2%) hydrogen and 60% helium. H2/He mixed fuel is recommended over H2/N2 fuel because of better linearity in concentration out- put. Such blends are supplied by many gas vendors specifically for this use, with a guaranteed maximum total hydrocarbon content of 0.5 ppm, measured as methane. This specification should be used when obtaining these mixtures.
Note
The fuel restrictor is marked with a red dot, and the sample capillary is marked with a red or green dot for mixed fuel applications.
d.Burner Air Specification
In order to ensure a low background signal, burner air should contain less than 1 ppm maximum total hydrocarbon content. An alternate source for burner air and zero gas (see Section
e.Purge Air Specification
Instrument quality air or nitrogen is re- quired for the safety purge system.
f.Regulated Air Specification
Instrument quality air or nitrogen is re- quired. The air should contain less than 2 ppm maximum total hydrocarbon con- tent.
g.Calibration Gases Specification
Calibration method and gases depend on the operating range, and the desired measurement accuracy. In all methods, zero and span gases are used, and are
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |