Model NGA2000 HFID
Instruction Manual
WARNING.
HIGH TEMPERATURE
This equipment is used in the analysis of sample gases at temperatures of up to 250°C. All components and material in contact with the sample, the oven and the burner can reach this temperature level.
Operate this equipment only when covers are secured. Servicing requires access to "hot" parts which can cause serious injury. Refer servicing to qualified personnel.
NOTE
This Analyzer Module is completely
WARNING
PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only
CAUTION
PURGE AIR REQUIREMENTS
This Analyzer Module must be used in conjunction with a device (Control Module or PC Interface) that can actively monitor network variables related to pressure or flow of the continuous dilution purge, or the front panel LEDs of the Analyzer Module, as installed, must be visible. The purpose of this requirement is to maintain adherence to ANSI/NFPA 496 standard which assures the continued viability of the purge system. Under no circumstances should any pressure or flow indicator be connected to the PURGE AIR OUT outlet of the Analyzer Module because this may affect the sealing performance of the module.
CAUTION
PRESSURIZED GAS
This module requires calibration with a known standard gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders on page
Rosemount Analytical Inc. A Division of Emerson Process Management | Preface |