3 Specifications and EnvironmentalConditions

3.1 Specifications

ResistanceRange 0
Ω
to400
Ω
,auto-ranging
ResistanceAccuracy, one year
1
0
Ω
to20
Ω
:0.0005
Ω
20
Ω
to400
Ω
:0.0025% (25 ppm) of reading
ResistanceAccuracy, short
term
1,2
0
Ω
to30
Ω
:0.0005
Ω
30
Ω
to400
Ω
:0.0015% (15 ppm) of reading
TemperatureRange
3
–200°Cto 962°C (–328°F to 1764°F)
TemperatureAccuracy
1,3, 4
–100°C:0.004°C
0°C:0.006°C
100°C:0.009°C
200°C:0.012°C
300°C:0.015°C
400°C:0.018°C
500°C:0.021°C
600°C:0.024°C
Temperature Coefficient of
Resistance
1
1ppm/°C
ResistanceResolution 0
Ω
to20
Ω
:0.0001
Ω
20
Ω
to400
Ω
:0.001
Ω
TemperatureResolution 0.001°C
Probe NominalR(0.01°C): 25
Ω
to100
Ω
RTD,PRT, or SPRT
ProbeConnection 4-wirewith shield, 5-pin DIN connector
Maximumacceptable lead
resistance
100
Ω
ProbeCharacterizations ITS-90sub-ranges 4, 6, 7, 8, 9, 10, and 11
IPTS-68:R
0
,
α
,
δ
,a
4
,and c
4
Callendar-VanDusen: R
0
,
α,δ
,and
β
ProbeExcitation Current 0.5and1 mA , user selectable, 2Hz
MeasurementPeriod 1second
DigitalFilte r Exponential,0 to 60 seconds time constant (user-selectable)
Communications RS-232serial standard
IEEE-488(GPIB) optional, conforms to IEEE-488.1, capability AH1,
SH1,T6, L4, DC1
Display 8-digit,7-segment, yellow-green LED; 0.5 inch high characters
Clockaccuracy, typical 0.01%
OperatingTemperature Range Fullaccuracy: 16°C to 30°C
Absolute: 0°C to 55°C
ACpower 115VAC ±10%, 50/60 Hz, 10 W, nominal, 1 A maximum
230 VAC ±10%, 50/60 Hz, 10 W, 1 A, nominal (optional)
Detachablepower cord
Size 5.6inches (14.3 cm) wide x 7.1 inches (18.1 cm) deep x 2.4 inches
(6.1cm) h igh

9

3 Specifications and Environmental Conditions

Specifications