277 Digital Communications InterfaceRemote Commands
Command Description
MeasurementCommands
T readmeasurement (includes label, unit, and time)
F[ETCH?] readmeasurement value (SCPI compatible)
SA[=[[[hh:]mm:]ss] read[or set] serial sample period
U=C|F|K|O selectunits
ST[=ON/OF] read[or set] the time stamp
CL[=hh:mm:ss] read[or set] the system clock
ProbeCharacterization Commands
PR[=90|68|R|S] read[or select] the characterization type
R0[=<value>] read[or set] R0 or R(0.01°C) depending on the selected characterization
AL[=<value>] read[or set]
α
forthe Callendar-Van Dusen or IPTS-68 characterization
DE[=<value>] read[or set]
δ
forthe Callendar-Van Dusen or IPTS-68 characterization
BE[=<value>] red[or set]
β
forthe Callendar-Van Dusen characterization
A4[=<value>] read [or set]
a
4
forthe ITS-90 or IPTS-68 characterization
B4[=<value>] read [or set]
b
4
forthe ITS-90 characterization
C4[=<value>] read[or set]
c
4
forthe IPTS -68 characterization
A
n
[=<value>] read [or set]
a
6
,a
7
,a
8
,a
9
,a
10
,or
a
11
forthe ITS-90 characterization
B
n
[=<value>] read [or set]
b
6
,b
7
,b
8
,
or
b
9
forthe ITS-90 characterization
C
n
[=<value>] read[or set]
c
6
or
c
7
forthe ITS-90 characterization
D6[=<value>] read[or set]
d
6
forthe ITS-90 characterization
SC[=68/90] read[or set] IPTS-68 scale conversion
CO=<value> testresistance to temperature conversion
SampleParameter Commands
FI[=<value>] read[or set] filter time constant
CU[=<value>] read[or set] probe current
PS[=<value.] read [or set] the power saver period
Communicationparameter commands
DU[=F/H] read [or set] serial sample duplex mode
LF[=ON/OF] read [or set] serial linefeed
Table 6 Command List