Example1:
A PRTwas calibrated to ITS-90 and its calibration certificate states values for
coefficientsRtpw, a4, b4, a8, and b8. Set the 1502A parameters with values
from the certificateas follows.
Example2:
A PRTwas calibrated to ITS-90 and its calibration certificate states values for
coefficientsRtpw, a5 and b5. Set the 1502A parameters with values from the
certificateas fo llows:
19
6 General Operation
Selecting the Probe Characterization
1502ACoefficient CertificateValue
R0.01 Rtpw
Aa8
bb8
C0
d0
A4 a4
b4 b4
Table 3 Setting CoefficientsRtpw, a8, b8, a4, and b4
1502ACoefficient CertificateValue
R0.01 Rtpw
Aa5
bb5
C0
d0
A4 a5
b4 b5
Table 4 Setting CoefficientsRtpw, a5, and b5