to pins4 and 5. (Pins 1 and 5 source current and pins 2 and 4 sense the poten-
tial.)If a shield wire is present it should be connected to pin 3.
A two-wireprobe can also be used with the 1502A. It is connected by attaching
one wireto both pins 1 and 2 of the plug and the other wire to both pins 4 and
5. If a shieldwire is present it should be connected to pin 3. Accuracy may be
significantlydegraded using a two-wire connection because of lead resistance.
1502A Thermometer Readout
User’s Guide
12
1
24
5
RTDSensor
ProbeConnector
3
Shield
Figure1 Connecting a four-wire probe