LF=OF<EOS> disables linefeed
7.3.5 Calibration CommandsThe following commands are used in calibrating the instrument.
7.3.5.1 Entering the Password
In order to setthe calibration parameters the password must be issued first. The
followingcommand enables access to the calibration parameters:
*PA=2051<EOS> enables the calibration commands
Calibration parameters can be lockedout again by sending *PA=0 or by cycling
the power.
7.3.5.2 Setting the Menu Lockout
The followingcommands can be used to select the menu lockout options:
*LO=CA<EOS> locks out only the calibration menu
*LO=AL<EOS> locks out all menus
7.3.5.3 Setting the Calibration Coefficients
The instrument calibration coefficients are used to maintain the resistance mea-
surement accuracyof the 1502A. These coefficients must not be changed ex-
cept by aqualified technician during the calibration of the 1502A. The
following commands can be used to set the instrument calibration coefficients:
*C0=<value><EOS> sets the calibration parameter CAL0
*C1=<value> <EOS> sets the calibration parameter CAL100
*C4=<value><EOS> sets the calibration parameter CAL400
7.3.5.4 Setting the Serial Number
The followingcommand is used to set the serial number of the 1502A:
*SN=<value><EOS> sets the instrument’s serial number
7.3.6 Other CommandsRemaining commands are described below.
7.3.6.1 Instrument Identification
The followingcommand returns the model number and firmware version
number:
*VER<EOS> returns the model and firmware version numbers
The syntax of the response isas follows:
1502A Thermometer Readout
User’s Guide
32