8 Calibration Procedure
The 1502A uses a three-pointcalibration scheme with a quadratic polynomial
correction function to maintain the accuracyof its resistance measurement. The
three calibration points are at 0Ω, 100Ω, and 400Ω.Three calibration pa-
rameters determine the correction function:CAL0, CAL100, and
CAL400.The CAL0 parameter sets the correction at 0Ωresistance (but
does not affectthe correction at 100Ω).The CAL100 parameter sets the
correction at 100Ωresistance (but does not affectthe correction at 0Ω).
The CAL400 parameter sets the correction at 400Ωresistance (but does
not affectthe correction at 0 and 100Ω). Adjusting the calibration param-
eters directly affects themeasurem entat the specific resistances.For
example,increasing the CAL100 parameter by 0.005 increases the
measured value at 100Ωby0.005Ω.
8.1 Accessing the Calibration Parameters
The calibration parameters are accessed inthe Cal menu. Th e calibration pa-
rameters are protected by requiring thecorrect password to access them. Press
the Menu button,“SEt?” appears. Press the Ω/Exit button and hold it down
for one second, “CAL” appears briefly.The display will show “PA= 0000”and
allowyou to change the number to the correct password. You must enter the
password(“2051”). Usethe Land Rbuttons to move between the password
digitsand the Uand Dbuttons to increase or decrease the value of a digit.
Press Enter when all the digits are correct. If the password is entered correctly
the firstparameter in the calibratio n menu will appear.
The firstparameter inthe Cal menu is the lockout control parameter, indicated
on the display as “LOCOUt”. This parameter has two options, “CAL”and
“ALL”. “CAL” (default) locks out the calibration menu only. “ALL” locks out
all menusand access to any menu requires the correct password. Use the L
and Rbuttonsto select the lockout option and press Enter to continue. The
instrument calibration parameters follow.
The calibration parameters appear with thename shown briefly then the value.
Youcan change the sign and digits of each parameter. Use the Land Rbut-
tons tomove between digits and the Uand Dbuttons to increase or decrease
the valueo f the digit. Press Enter to saveth e newvalue.
The last parameter in the Cal menu is the factoryr esetfunction. This can be
used to clear the internalmemory and reset all parameters to the defalut values.
This isintended to be used only at the factory.
The calibrationparameters can also be set using remote commands through the
RS-232 or IEEE-488 interface.The *PA=<password><EOS> command must
be used first,using the correct password (“2051”), to enable access to the cali-
bration parameters. Lockoutprotection is automatically set by cycling the
35
8 Calibration Procedure
Accessing the Calibration Parameters