10.1.1.1 Immunity Testing
The instrumentwas tested to the requirements for industrial locations. This al-
lowsthe instrument to be used in all types of locations from the laboratory to
the factoryfloor. Criterion C was used for Electrostatic Discharge (ESD, IEC
61000-4-2) and ElectricFast Transit (EFT, Burst, IEC 61000-4-4). If the instru-
ment is subjectedto EFT conditionsat 2kV, the instrument may require the user
to cyclethe power to return to normal operation.
10.1.1.2 Emission Testing
The instrument fulfills the limit requirements for Class A equipment but does
not fulfillthe limit requirements for Class B equipment. The instrument was
not designedto be used in domestic establishments.
10.1.2 Low Voltage Directive (Safety)
In order to complywith the European Low Voltage Directive (73/23/EEC),
Hart Scientificequipment has been designed to meet the IEC 1010-1 (EN
61010-1) and theIEC 1010-2-010 (EN 61010-2-010) standards.