￿ Open the SCOPE INPUTS menu.

Highlight PROBE AC ADJUST.

Highlight ADJUST 10:1 PROBE.Adjust the trimmer screw in the

probe housing to give an optimum square wave.

Manual background

Open the SCOPE INPUTS menu.

Open the PROBES submenu.

Highlight PROBE AC ADJUST.

Open the PROBE AC ADJUST submenu.

Highlight ADJUST 10:1 PROBE.

A square wave appears on the screen.

Manual background

Manual background

Adjust the trimmer screw in the
probe housing to give an optimum square wave.

Return to normal mode.