Fluke 123

Users Manual

Multimeter disturbance:

VDC, VAC, and VAC+DC with STL120 and short ground lead.

OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.

Table 3

Disturbance less than 1% of full scale

E= 3 V/m

E= 10 V/m

Frequency range 10 kHz to 27 MHz

VDC, VAC, VAC+DC

OHM, CONT, DIODE CAP

Frequency range 27 MHz to 1 GHz

VDC, VAC, VAC+DC

OHM, CONT, DIODE CAP

500 mV to 1250V

500Ω to 30 MΩ

50 nF to 500 μF

500 mV to 1250V

500Ω to 30 MΩ

50 nF to 500 μF

500 mV to 1250V

500Ω to 30 MΩ

50 nF to 500 μF

500 mV to 1250V

500Ω to 30 MΩ

50 nF to 500 μF

Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.