Fluke fluke123 Advanced Meter Functions, Continuity CONT, Diode, Zero Set, Fast/Normal/Smooth

Models: fluke123

1 77
Download 77 pages 58.02 Kb
Page 68
Image 68
Continuity (CONT)

Specifications

4

Dual Input Meter

Measurement Current

.......................... 0.5 mA to 50 nA

decreases with increasing ranges

Open Circuit Voltage

<4V

Continuity (CONT)

<(30Ω ±5Ω) in 50Ω range

Beep

Measurement Current

0.5 mA

Detection of shorts of

.......................................... 1 ms

Diode

Advanced Meter Functions

Zero Set

Set actual value to reference

Fast/Normal/Smooth

Meter settling time Fast: 1s @ 1 μs to 10 ms/div.

Meter settling time Normal: 2s @ 1 μs to 10 ms/div.

Meter settling time Smooth: 10s @ 1 μs to 10 ms/div.

Measurement Voltage:

 

@0.5 mA

>2.8V

@open circuit

<4V

Accuracy

±(2% +5 counts)

Measurement Current

0.5 mA

Polarity

+ on input A, - on COM

Capacitance (CAP)

50 nF, 500 nF, 5 μF, 50 μF, 500 μF

Ranges

Accuracy

±(2% +10 counts)

Full Scale Reading

....................................

5000 counts

Measurement Current

5 μA to 0.5 mA

increases with increasing ranges

Dual slope integrating measurement with parasitic serial and parallel resistance cancellation.

Touch Hold (on A)

Captures and freezes a stable measurement result. Beeps when stable. Touch Hold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals.

TrendPlot

Graphs meter readings of the Min and Max values from 15 s/div (120 seconds) to 2 days/div (16 days) with time and date stamp. Automatic vertical scaling and time compression.

Displays the actual and Minimum, Maximum, or average (AVG) reading.

Fixed Decimal Point

Possible by using attenuation keys.