Fluke fluke123 Dual Input Meter, Input A and Input B, DC Voltage VDC, Peak, Frequency Hz

Models: fluke123

1 77
Download 77 pages 58.02 Kb
Page 66
Image 66
Dual Input Meter

Specifications

4

Dual Input Meter

Dual Input Meter

The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C.

Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1%. More than one waveform period must be visible on the sceen.

Input A and Input B

DC Voltage (VDC)

 

 

Ranges

500 mV, 5V, 50V, 500V, 1250V

Accuracy

±(0.5% +5 counts)

Normal Mode Rejection (SMR)

>60 dB

 

 

@ 50 or 60 Hz ±1%

Common Mode Rejection (CMRR)

>100 dB @ DC

 

>60 dB @ 50, 60, or 400 Hz

Full Scale Reading

5000 counts

True RMS Voltages (VAC and VAC+DC)

Ranges

500 mV, 5V, 50V, 500V, 1250V

Accuracy for 5 to 100% of range

 

DC coupled:

...............

±(1% +10 counts)

DC to 60 Hz (VAC+DC)

1 Hz to 60 Hz (VAC)

±(1% +10 counts)

AC or DC coupled:

±(2.5% +15 counts)

60 Hz to 20 kHz

 

20 kHz to 1 MHz

±(5% +20 counts)

1 MHz to 5 MHz

±(10% +25 counts)

5 MHz to 12.5 MHz

±(30% +25 counts)

5 MHz to 20 MHz (excl. test leads or probes)

 

 

±(30% +25 counts)

AC coupled with 1:1 (shielded) test leads

60 Hz (6 Hz with 10:1 probe)

-1.5%

50 Hz (5 Hz with 10:1 probe)

-2%

33 Hz (3.3 Hz with 10:1 probe)

-5%

10 Hz (1 Hz with 10:1 probe)

-30%

DC Rejection (only VAC)

>50 dB

Common Mode Rejection (CMRR)

>100 dB @ DC

 

>60 dB @ 50, 60, or 400 Hz

Full Scale Reading

5000 counts

The reading is independent of any signal crest factor.

Peak

 

 

Modes

Max peak, Min peak, or pk-to-pk

Ranges

500 mV, 5V, 50V, 500V, 1250V

Accuracy:

 

 

Max peak or Min peak

5% of full scale

Peak-to-Peak

10% of full scale

Full Scale Reading

500 counts

Frequency (Hz)

 

 

Ranges

1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz,

100 kHz,1 MHz, 10 MHz, and 40 MHz Frequency Range in Continuous Autoset .....................

15 Hz (1 Hz) to 30 MHz

59

Page 66
Image 66
Fluke fluke123 Dual Input Meter, Input A and Input B, DC Voltage VDC, True RMS Voltages VAC and VAC+DC, Peak, Frequency Hz