Duty Cycle (DUTY)

Fluke 123

Users Manual

Accuracy:

±(0.5% +2 counts)

@1 Hz to 1 MHz

@1 MHz to 10 MHz

±(1.0% +2 counts)

@10 MHz to 40 MHz

±(2.5% +2 counts)

Full Scale Reading

10 000 counts

Duty Cycle (DUTY)

 

Range

2% to 98%

Frequency Range in Continuous Autoset

 

15 Hz (1 Hz) to 30 MHz

Accuracy:

±(0.5% +2 counts)

@1Hz to 1 MHz

@1 MHz to 10 MHz

±(1.0% +2 counts)

@10 MHz to 40 MHz

±(2.5% +2 counts)

Resolution

0.1%

Pulse Width (PULSE)

Frequency Range in Continuous Autoset ......................

 

 

15 Hz (1 Hz) to 30 MHz

Accuracy:

............................

±(0.5% +2 counts)

@1 Hz to 1 MHz

@1 MHz to 10 MHz

±(1.0% +2 counts)

@10 MHz to 40 MHz

±(2.5% +2 counts)

Full Scale Reading

1000 counts

Amperes (AMP)

with optional current probe

Ranges

same as VDC, VAC, VAC+DC, or PEAK

Scale Factor ...

1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A

Accuracy

same as VDC, VAC, VAC+DC, or PEAK

 

 

(add current probe uncertainty)

Temperature (TEMP).... with optional temperature probe

Range

200 °C/div (200 °F/div)

Scale Factor

1 mV/°C and 1 mV/°F

Accuracy

as VDC (add temp. probe uncertainty)

Decibel (dB)

 

 

0 dBV

1V

0 dBm (600Ω /50Ω)

1 mW

 

 

referenced to 600Ω or 50Ω

dB on

VDC, VAC, or VAC+DC

Full Scale Reading

1000 counts

Crest Factor (CREST)

 

Range

1 to 10

Accuracy

±(5% +1 count)

Full Scale Reading

100 counts

Phase

 

 

Modes

A to B, B to A

Range

0 to 359 degrees

Accuracy

±(1 degree +1 count)

Resolution

1 degree

Input A

Ohm (Ω)

Ranges .......500Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ

Accuracy:

±(0.6% +5 counts)

Full Scale Reading:

 

500Ω to 5 MΩ

5000 counts

30 MΩ

3000 counts