Fluke fluke123 Dual Input Oscilloscope, Vertical, Frequency Response, Rise Time, Input Impedance

Models: fluke123

1 77
Download 77 pages 58.02 Kb
Page 63
Image 63
Dual Input Oscilloscope

Fluke 123

Users Manual

Dual Input Oscilloscope

Vertical

Frequency Response

DC Coupled:

excluding probes and test leads:

...................................

 

DC to 20 MHz (-3 dB)

with STL120 1:1 shielded test leads:

DC to 12.5

MHz (-3 dB)

 

DC to 20

MHz (-6 dB)

with PM8918 10:1 probe (optional accessory):

 

DC to 20

MHz (-3 dB)

AC Coupled (LF roll off):

 

 

excluding probes and test leads

<10 Hz (-3 dB)

with STL120

<10 Hz (-3dB)

with PM8918

<1 Hz (-3 dB)

Rise Time

 

 

excluding probes and test leads

<17.5 ns

Input Impedance

 

excluding probes and test leads

................ 1 MΩ//12 pF

with BB120

1 MΩ//20 pF

with STL120

1 MΩ//225 pF

with PM8918

10 MΩ//15 pF

Sensitivity

5 mV to 500 V/div

Display Modes

A, -A, B, -B

Max. Input Voltage A and B

 

direct or with test leads

600 Vrms

with BB120

300 Vrms

(For detailed specifications, see “Safety”)

Max. Floating Voltage

 

from any terminal to ground

600 Vrms

 

up to 400 Hz

Resolution

8 bit

Vertical Accuracy

±(1% + 0.05 range/div)

Max. Vertical Move

±4 divisions