Fluke Food Safety Thermometer Hot Temperature Verification Check, Field Verification of Accuracy

Models: Food Safety Thermometer

1 22
Download 22 pages 57.22 Kb
Page 15
Image 15

Noncontact Food Safety Thermometer

Field Verification of Accuracy

0ºC

(32ºF)

ebq11.eps

Figure 8. Cold Temperature Verification Check

The noncontact (IRT) measurement should be within ±1 °C (±2 °F) of the reference probe reading (nominally 0 °C (32 °F)).

The probe temperature reading should be within ±0.5 °C (±1 °F) of the reference probe reading.

Hot Temperature Verification Check

1.Follow the same procedure as above, substituting hot water (>140 °F/60 °C). Hot tap water is adequate for the procedure.

2.Repeat steps two and three from above.

Note

Due to evaporative cooling on the surface of the hot water, it is particularly important that the surface of the water be continually stirred while making the IR measurement (Figure 9).

11

Page 15
Image 15
Fluke Food Safety Thermometer user manual Hot Temperature Verification Check, Field Verification of Accuracy