Fluke Series II user manual 7-12, OneTouch Series, Users Manual, Stop at Upstream Incr by, 10 Mb

Models: Series II

1 127
Download 127 pages 19.34 Kb
Page 95
Image 95

OneTouch Series II

Users Manual

Stop at

Upstream

Incr by

Select the highest speed for the range of downstream rates to be tested.

Select the speed of the upstream traffic generated during the test.

Select the increment size of the downstream speeds within the range you defined with the Start at and Stop at selections. The test increments the downstream speed and continues testing only if the channel passes 95% of the transmitted data at the current speed.

qUser 1 and User 2: Allows you to define your own upstream and downstream rates.

The total upstream and downstream rates must not exceed the values listed below:

10 Mb

Frame Size

Max Bps

 

64

37Mbps

128

64Mbps

256

78Mbps

512

87Mbps

768

87Mbps

1024

91Mbps

1280

93Mbps

1518

85Mbps

7-12

Page 95
Image 95
Fluke Series II user manual 7-12, OneTouch Series, Users Manual, Stop at Upstream Incr by, 10 Mb