Accessory Assembly - Head & Foot Half Rails

Head-End Half Rail KIT: ZA78400

(Includes part 999-0784-908G —Left side and

part 999-0784-907G - right side, plus hardware bag)

HeAd

right

Left

Foot

HEAD

BOARD

note: release trigger is located at this end.

LeFt HALF rAIL

999-0784-908G sHoWn

Item #

Qty.

Part #

Description

 

 

 

 

 

 

 

Head-End Half Rail: Left-Hand Side - 76”

1

1

999-0784-908G

Head-End Half Rail: Left-Hand Side - 80”

 

 

 

 

 

 

 

Foot-End Half Rail: Right-Hand Side - 80” Only (Not shown)

 

 

 

 

1

1

999-0794-908G

Foot-End Half Rail: Right-Hand Side - 76” Only (Not shown)

 

 

 

 

 

 

 

Foot-End Half Rail: Left-Hand Side - 80” Only (Pg. 18)

 

 

 

 

1

1

999-0784-907G

Head-End Half Rail: Right-Hand Side - 76” (Not shown)

 

 

 

 

 

 

 

Head-End Half Rail: Right-Hand Side - 80” (Not shown)

 

 

 

 

1

1

999-0794-907G

Foot-End Half Rail: Left Hand Side - 76” Only (Pg. 18)

 

 

 

 

 

 

 

Clevis Pin - 16mm. Dia. x 105mm. w/cross-hole,

2

2

999-0784-100

circle cotter, lanyard cable (LBK-3-06)

 

 

 

and Hair Pin Clip

 

 

 

 

1.All Half Rails (#1) mount to the bed in a similar fashion. Left or right designations are based on the patient’s orientation as he/she lies in the bed.

2.Insert the two Clevis Pins, with Circle Cotter and Lanyard Cable (#2), into the two Half Rail arms as shown.

3.Attach the Half Rail to the bed frame brace by aligning the arms and Clevis Pins to the holes on the brace. Slide the Clevis Pins so that they extend past the brace on the underside of the bed.

4.Insert the two Hair Pin Clips (#3) into the holes in the Clevis Pins to secure the rail to the bed.

!WARNING: The mattress support platform should always be placed in the flat position when not attended.

5.IMPORTANT: Lubricate all pivot joints as needed. Aerosol White Lithium Grease is recommended.

When assessing the Risk for Entrapment, you need to consider your bed, mattress, head- and footboards, half rails, assist bars, and other accessories as an entire system. All bed systems are evaluated for compliance to the FDA/CDRH “Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment” guidelines.

6.TO OPERATE: Depress the Release Trigger to lower or raise the half rail assembly.

A resident/patient’s condition could lead to resident/patient entrapment. It is extremely important to review the resident/patient’s physical and mental condition, and initiate an appropriate individual care plan to address any entrapment risks.

NO SPECIAL Tools REQUIRED

Basic American Medical products, a division of GF Health products, Inc. • Matrix II • rev d Manual

p17

Page 20
Image 20
Graham Field Mp717404, Mp7074, Mp7114, MM7114 manual Ccessory Assembly Head & Foot Half Rails, Head-End Half Rail KIT ZA78400