The following calibration procedure can be used when the remote scale display is not present. The Multi- Function Button (see page 44) is used to advance to the next stage of calibration and the LED display notifies the operator which ‘calibration stage’ is active.
1. Temporarily remove the platter and place it in a safe location.
Note: It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures statutes and regulations applicable in your area by checking with the appropriate government agency before placing a newly calibrated unit into service or removing any official seals.
2.If this is a currently installed scanner/Diva scale, there will be either an adhesive label seal or wire sealed cover over the scale calibration button. New installation units will only have a
To remove the adhesive label seal, peel off the label, and discard.
To remove the wire seal, cut the wire, and discard. Fully loosen the security screw and lift off the cover. Place the security screw and cover in a safe location.
Figure 46. Diva Scale security seals
* These procedures are for MS2420/MS2430 Scanner/Diva Scale models only.