Table 1 Non-queued ACI commands

 

 

ACI Command

Recommended time-out value

Get Drive Info

5s

Get Drive Status

5s

 

 

Get Drive Configuration

5s

 

 

Get Error Info

5s

 

 

Get Buffer Size

5s

 

 

No Op

5s

 

 

Acknowledge Attention

5s

Table 2 Queued ACI commands

 

 

 

ACI Command

Recommended time-out value

ACI Load—immediate

5s

ACI Load—non-immediate (drive idle, unloaded)

300s

 

 

ACI Unload—immediate

5s

 

 

ACI Unload—non-immediate (tape loaded, at EOM,

300s or 9000s depending on

drive idle)

implementation strategy

 

 

Set Drive Configuration (tape loaded, at EOM, SCSI

300s or 9000s depending on

unload)

implementation strategy

 

 

ACI Reset—ACI bus

5s

 

 

ACI Reset—drive

5s

 

 

Set Baud Rate

5s

Treatment of reserved fields

To ensure forwards compatibility with future versions of the ACI, automation controller firmware should set any command fields labelled as ‘Reserved’ to zero. Likewise, automation controller firmware should mask off any response fields labelled as ‘Reserved’ during the processing of tape drive responses. This will allow older versions of automation controller firmware to operate successfully with newer versions of tape drive firmware.

Recommended power-up sequence

After power-up, HP recommends that the automation controller wait until it has received at least one ASCII <ENQ> character from the tape drive before attempting a command-response transaction. HP Ultrium tape drives use a two-step power-up sequence and the drive sends <ENQ> to signal the transition between the steps. The drive sends the first <ENQ> within 500 milliseconds of exiting the reset state after receiving power. Note that the default baud rate at power-up is 9600.

Consider sending Get Drive Info as the first command, either packetized or primitive. This retrieves a variety of useful identifying information identifying the tape drive, including the version of the ACI protocol that the tape drive supports.

HP LTO Ultrium 4 drives technical reference manual, volume 2: software integration 43 HP restricted