Invacare SP Safety Inspection/Troubleshooting, Possible Solution, Symptom, Probable Cause, Number

Models: SP

1 72
Download 72 pages 50.43 Kb
Page 32
Image 32
NUMBER

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Service Indicator Light Diagnostics

NUMBER

ERROR CODE

 

OF

POSSIBLE SOLUTION

DESCRIPTION

FLASHES

 

 

 

 

 

 

1

User Fault

Release joystick to neutral and try again.

 

 

 

2

Battery Fault

Charge the batteries. Refer to Charging Batteries on page 62.

 

 

Check that battery cables are connected properly. If necessary,

 

 

replace batteries. Refer to Replacing Batteries on page 55.

 

 

 

3

Left Motor Fault

Contact Invacare/Dealer for service.

 

 

 

4

Right Motor Fault

Contact Invacare/Dealer for service.

 

 

 

5

Left Park Brake Fault

Contact Invacare/Dealer for service.

 

 

 

6

Right Park Brake Fault

Contact Invacare/Dealer for service.

 

 

 

7

Remote Fault

Check to make sure joystick is connected properly.

 

 

Contact Invacare/Dealer for service.

 

 

 

8

Controller Fault

Contact Invacare/Dealer for service.

 

 

 

9

Communications Fault

Contact Invacare/Dealer for service.

 

 

 

10

General Fault

Contact Invacare/Dealer for service.

 

 

 

11

Incompatible or incorrect

Wrong type of remote connected. Contact Invacare/Dealer

 

Remote

for service.

 

 

 

MPJ™+, PSR+, PSF+ Joysticks or Displays

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

SPM L Park Brake Fault or

Motor lock levers disengaged

Engage motor lock levers. Refer to

SPM R Park Brake Fault displays

(Error code E9 or E10).

Disengaging/Engaging Motor Lock

 

Levers on page 48.

and wheelchair does not drive.

 

 

 

 

 

 

CHARGER PLUGGED IN displays.

Battery charger connected

Unplug battery charger from the

 

(Error code E28).

wheelchair. Refer to Charging Batteries

 

 

on page 62.

 

 

 

SPM Battery Fault displays and the

Batteries need to be charged

Charge batteries. Refer to Charging

wheelchair does not drive.

(Error code E14).

Batteries on page 62. If batteries fail to

 

charge properly, check battery charger

 

 

 

 

or replace batteries. Refer to Replacing

 

 

Batteries on page 55.

 

 

 

JOYSTICK TIMEOUT displays and

Joystick or input device is

Turn off power, reconnect the joystick

the wheelchair does not drive.

disconnected (Error code 32).

of input device and turn power on.

 

 

 

 

 

JS REV TOO LARGE

The joystick or input device is

Replace joystick or input device.

JS FWD TOO LARGE

sending a value outside of the

 

reverse, forward, left or right

 

JS LFT TOO LARGE or

 

limits (Error codes E01, E02,

 

JS RGT TOO LARGE

E03 or E04).

 

displays and the wheelchair does not

 

 

drive.

 

 

 

 

 

TDX™ SP

32

Part No 1143190

Page 32
Image 32
Invacare SP Safety Inspection/Troubleshooting, Possible Solution, Symptom, Probable Cause, Solutions, Number, Description