SECTION 5—TYPICAL PRODUCT PARAMETERS

Oxygen Concentration:*

87% to 95.6%, after initial warm-up

*Based on an atmospheric

period (approximately 5 minutes)

pressure of 14.7 psi (101 kPa) at

 

70°F (21°C)

 

 

 

Oxygen Flow:

Pulse flow delivery. Bolus Volume Ranges

 

From 300 - 840cc settings 1 to 5.

 

 

Dimensions:

10” high x 7” wide x 4” deep (25.4 cm

 

high x 17.8 cm wide x 10.2 cm wide)

 

 

Weight:

6.0 lbs (2.72 kg); 1.3 lbs (0.68 kg) for

 

supplemental battery

 

 

Battery Duration (times are

Setting 1 = 3.5 hrs

approximate):

 

Setting 2 = 2.5 hrs

 

 

 

 

Setting 3 = 2.0 hrs

 

 

 

Setting 4 = 1.5 hrs

 

 

 

Setting 5 = 1.0 hrs

 

 

Battery Recharge Time:

4 hours

NOTE: Recharge time increases if

 

battery is charging while unit is

 

running.

 

 

 

Humidity

Operating Humidity: 15% to 60% non

 

condensing

 

 

 

Storage Humidity: up to 95% non

 

condensing

 

 

Temperature Range:

Operating temperature: 41°F to 95°F

(All power sources)

(5°C to 35°C)

 

 

 

Storage temperature: -2°F to 140°F

 

(-20°C to 60°C)

 

 

Extended Temperature Range:

Operating Temperature:

(Using AC or DC adapters)

95°F to 104° (35°C to 40°C)

 

Continuous use - all settings

 

 

Extended Temperature Range

Operating Temperature:

(Using Internal Battery)

95°F to 104° (35°C to 40°C)

 

Continuous use - settings 1, 2 and 3

 

45 minutes (max) - setting 4

 

30 minutes (max) - setting 5

 

 

Part No 1148112

15

Invacare XPO2

Page 15
Image 15
Invacare XPO100B user manual Battery is charging while unit is Running