SECTION 6—OPERATING INSTRUCTIONS
FLOW SETTING | AVERAGE PULSE VOLUME (AT 20 | |
|
| BREATHS PER MINUTE) |
|
|
|
1 |
| 15 ml |
|
|
|
2 |
| 23 ml |
|
|
|
3 |
| 31 ml |
|
|
|
4 |
| 37 ml |
|
|
|
5 |
| 42 ml |
|
|
|
Indicator Lights | On/Off Button |
Flow Selection
FIGURE 6.9 Adjusting the Pulse Flow Setting
Using the XPO2 Portable Concentrator
1.Turn the XPO2 On by pressing the On/Off button.
2.Breathe normally through your nose. Breathing through your mouth may result in less effective oxygen therapy. The XPO2 will put out a measured pulse of oxygen each time it detects inhalation. The blue flow setting indicator light will blink each time an inhalation is detected and the XPO2 outputs a pulse of oxygen.
3.Turn the XPO2 Off by pressing and holding the power On/Off button for approximately two seconds.
NOTE: The time required for the XPO2 to reach maximum oxygen concentration after being turned On is approximately 5 minutes.
Invacare XPO2™ | 32 | Part No 1148112 |