SECTION
ALARM/WARNING: |
| |
SYMPTOM | PROBABLE CAUSE | SOLUTION |
CHARGING INDICATOR |
| |
With the unit | Power source is no good, or | Try another power |
turned Off and | there is a loose connection. | outlet and check |
with the AC or |
| connections to charger |
DC charging |
| and unit. |
adapter plugged | Internal Battery is outside | Allow unit to cool down |
into the unit, no | the allowed temperature | to less than 95ºF, or |
Battery Gauge | range for charging. | warm up to 50ºF. |
segments are |
|
|
illuminated. |
|
|
Battery Gauge is |
|
|
dark. |
|
|
LOW BATTERY | WARNING: |
|
There is a single | Internal battery requires | Connect XPO2 to its AC |
audible beep and | charging. | or DC power adapter, |
the green 25% |
| or to its supplemental |
battery indicator is |
| battery pack. |
blinking. |
|
|
LOW BATTERY | ALARM: |
|
There is a double | Internal battery requires | 1. Connect XPO2 to its |
audible beep and | charging. | AC or DC power |
the green 25% |
| adapter, or to its |
battery indicator is |
| supplemental battery |
rapidly blinking. |
| pack. |
|
| 2. If no other power |
|
| source is available, |
|
| change to a reserve |
|
| oxygen supply. |
BATTERY DISCHARGED ALARM: |
| |
There is a triple | Internal battery is | 1. Connect XPO2 to its |
audible beep and | completely discharged and | AC or DC power |
the green 25% | needs to be recharged. | adapter, or to its |
battery indicator is |
| supplemental battery |
blinking very |
| pack. |
rapidly. |
| 2. If no other power |
|
| source is available, |
|
| change to a reserve |
|
| oxygen supply. |
Part No 1148112 | 41 | Invacare XPO2™ |