ALARM/WARNING: |
| |
SYMPTOM | PROBABLE CAUSE | SOLUTION |
BATTERY TOO | HOT/COLD ALARM: |
|
Continuous | Internal Battery has become | 1. Unplug chargers if |
audible beep and | too hot, or too cold, during | connected. |
red alarm | operation. | 2. Allow unit to cool |
indicator is |
| down to less than 95ºF, |
illuminated. |
| or warm up to 50ºF, |
|
| before resuming |
AND |
| charging or turning Off |
|
| the unit and turning the |
Flow indicators 1 |
| unit On again. |
& 4 are |
| 3. If the alarm continues, |
illuminated. |
| change to another |
|
| source of oxygen and |
|
| contact your equipment |
|
| provider. |
STUCK BUTTON | ALARM: |
|
Continuous | A stuck button is being | 1. Turn Off the unit |
audible beep and | detected upon power up. | 2. With the unit turned |
red alarm |
| Off, press each control |
indicator is |
| button looking for a |
illuminated. |
| stuck button |
|
| 3. Turn On the unit |
AND |
| while ensuring that you |
|
| are not continuing to |
Flow indicators 1 |
| hold the Power On/Off |
& 5 are |
| button once the lights |
illuminated. |
| begin to illuminate. |
|
| 4. If the alarm continues, |
|
| change to another |
|
| source of oxygen and |
|
| contact your equipment |
|
| provider |
Part No 1148112 | 43 | Invacare XPO2™ |