SECTION 7—CLEANING, CARE, AND MAINTENANCE
Invacare XPO2 38 Part No 1148112
Air Intake Filter
CAUTION
DONOToperatetheXPO2withouttheairintakefilter
installed.
NOTE:Forthisprocedure,refertoFIGURE7.1.
NOTE:Removetheairintakefilterandcleanitatleastonceeach
week.
1. Liftfiltercoverslightlyandpulldowntoremovetabsfrom
thegrooves.
2. Liftoutthefilter.
3. Useavacuumcleanerorwash withamildliquiddish
detergent(suchasDawn™)andwater.Rinsethoroughly.
4. Thoroughlydrythefilterandinspectforfraying,
crumbling,tearsandholes.Replacefilterifanyisfound.
5. Reinstallthefiltercoverbyplacingthetabsintheslotsand
engagingthemagnet.
NOTE:Use only Invacare part number 1148281 as the air intake
filter for yourXPO2.
FIGURE 7.1 Air Intake Filter
Cover
Filter
Magnet Cover Tabs
Magnet
Grooves