Invacare XPO100B user manual Maintenance, Radio Frequency Interference

Models: XPO100B XPO100

1 48
Download 48 pages 17.07 Kb
Page 9
Image 9

SECTION 1—GENERAL GUIDELINES

For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 5 minutes. Shorter periods of operation may reduce maximum product life.

The XPO2 is not designed to be used with a humidifier. Use of

this device with a humidifier may impair performance and/or damage the equipment.

When your automobile is turned Off, disconnect the car accessory power supply and remove the XPO2 from the

automobile. NEVER allow the XPO2 to be stored in a very hot

or cold automobile or in other similar, high or low, temperature environments. Refer to Typical Product Parameters on page 14.

DO NOT operate in temperatures below 41° F (5°C) or above 104° F (40° C) for extended periods of time.

The XPO2 MUST be used in an upright position. Invacare recommends the XPO2 not be used in the rain.

Maintenance

The XPO2 was specifically designed to minimize routine preventive maintenance. Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.

Radio Frequency Interference

This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601‐1‐2. These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation.

Part No 1148112

9

Invacare XPO2

Page 9
Image 9
Invacare XPO100B user manual Maintenance, Radio Frequency Interference