SECTION 8—TROUBLESHOOTING GUIDE

SECTION 8—TROUBLESHOOTING GUIDE

If your XPO2 Portable Oxygen Concentrator fails to operate properly, refer to the following chart for possible causes and solutions. If necessary, contact your service provider.

ALARM/WARNING:

 

SYMPTOM

PROBABLE CAUSE

SOLUTION

START-UP INDICATOR

 

XPO2 does not

On/Off button was not held

Try to power up the

operate when

down long enough.

unit again while

On/Off button

 

continuing to press the

pressed.

 

On/Off button until the

 

 

front panel lights begin

OR

 

to illuminate. This takes

 

 

typically 2-3 seconds.

Front panel lights

Internal battery is

Connect XPO2 to either

DO NOT

discharged (and

the AC or DC charger

illuminate upon

supplemental battery if

and retry.

start-up

connected to the unit).

 

 

Malfunction

Change to an alternate

OR

 

oxygen supply and

 

 

contact your equipment

All front panel

 

provider.

lights go dark.

 

 

NO BREATH DETECTED ALARM:

 

Continuous

XPO2 has not detected a

1. Verify the cannula is

audible beep and

breath for a predetermined

connected, not kinked,

red alarm

time period.

properly positioned and

indicator is on

 

you are breathing

continuously (not

 

through your nose.

blinking) upon

 

2. If the alarm continues,

start up.

 

change to another

 

 

source of oxygen and

 

 

contact your equipment

 

 

provider.

 

 

 

Invacare XPO2

40

Part No 1148112

Page 40