SECTION 6—OPERATING INSTRUCTIONS

Location

WARNING

NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.

Keep unit at least three inches away from walls, draperies, furniture, and the like.

Locate and position the XPO2 in a well ventilated space so that the air intake and the air exhausts are not obstructed.

Recommendations for Optimal Performance

Operating Temperature:

41°F to 95°F (5°C to 35°C)

 

 

Relative Humidity:

15% to 60%

 

 

Transport/Storage

-2°F to 140°F (-20°C to 60°C)

Temperature:

Allow unit to warm, or cool, to operating

 

temperature range before using.

 

 

Electrical:

No extension cords.

 

 

Altitude:

Up to 10,000 ft (3046 m) above sea level.

 

 

Tubing and Cannula:

4 ft cannula with a maximum 25 ft of crush

 

resistant tubing (DO NOT pinch).

 

 

Environment:

Smoke, pollutant and fume free. No

 

confined spaces (example: no closets).

 

 

Time of Operation:

Up to 24 hours per day when connected to

 

AC or DC power adapter.

 

 

Invacare XPO2