5. Assembly

Inserting the gas discharge lamps (Hg and Xe) into the 106z lamp housing

Hg and Xe lamps are powered by the separate ebq 100 supply unit.

Read the separate instruction manual provided with this supply unit.

The following gas discharge lamps may be used and require different lamp mounts (Fig. 19):

Type

Typical bulb life*

 

 

50 W high-pressure mercury burner (alternating current)

100 hrs.

100 W high-pressure mercury burner (direct current, stabilized/not stabilized

200 hrs.

100 W high-pressure mercury burner (direct current, stabilized/not stabilized, type 103 W/2)

300 hrs.

75 W High-pressure xenon burner (direct current, stabilized)

400 hrs.

* Please regard the data sheets of the burners.