5. Assembly
5.7 Equipping the Incident Light filter turret
The receptacles on the turret are numbered. According to your equipment, the individual filter and/or reflector cubes have already pre- assigned positions. A list is provided along with your shipment (“Identification Sheet”).
Insert the filter and reflector cubes in the follow- ing manner:
•Equip the incident light turret only when the microscope is switched off.
•Remove the face plate from the upper part of the microscope (Fig. 25). Turn the turret in any direction until the locking pin engages.
•Insert the filter or reflector cube into the mounting in front of you according to the identification sheet provided.
To do so, place the filter or reflector cube on the right side and press it to the left into the mounting (Fig. 26).
•Push the retention pin (25.2) and continue to turn the filter turret until you reach the next locking position.
Fig. 23 Filter cube | Fig. 24 Filter cube |
front side | back side |
Fig. 25 Removing the front panel
1Filter receptacle
2Retention pin
3Front panel
1
2
3
•Again make sure that the turret engages (retention pin unlocks) and insert the next filter and/or reflector cube as described above.
•When all filters and reflector cubes have been inserted, close the front cover plate again.
5.8 Polarizer and Analyzer
Fig. 26 Inserting the filter or reflector cubes
1Mounting
11
26