Lifescan Blood Glucose Monitor manual Control Solution Results

Models: Blood Glucose Monitor

1 89
Download 89 pages 5.76 Kb
Page 38
Image 38

Operation

Control Solution Results

When the control test result falls within the control range, the meter has passed the QC check and can be used to test patient blood.

If the result falls outside the expected range, repeat the test. Results that fall outside the expected range may indicate:

procedural error

control solution is expired, contaminated, beyond its discard date, or inadequately mixed

code number in meter does not match code number on test strip bottle

dirty meter or test strip holder

problem with test strip (expired, damaged, etc)

control solution testing temperature outside the range of 18°–30°C (64°–86°F)

too much or too little control solution applied to test strip